Microsoft LPF-00004 Manuel de l'opérateur Page 48

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 157
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 47
Page 30 PRINCIPLES OF OPERATION
C
YBER
A
MP
380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
Pre-filter Input-referred Input-referred
Gain Range Resolution
x1 ±3.00 V 100 µV
x10 ±300 mV 10 µV
x100 ±30.0 mV 1 µV
Rather than adding it to the input, the DC offset is added after the pre-filter amplifier for two reasons.
First, it is technically difficult to add an offset at the input without compromising either the noise, the
common-mode rejection ratio, or the input impedance. Second, by adding the DC level after the pre-
filter amplifier, better resolution is available for small signals. For example, if the user is recording
temperature from a semiconductor temperature probe putting out 10 mV/°C, an offset control with
100 µV resolution and ±3 V range corresponds to a temperature resolution and range of 0.01°C and
300°C respectively. This is more than sufficient for most applications. On the other hand, if the user
is recording from a blood-pressure strain gauge that puts out 25 µV/mmHg, an offset control with
100 µV resolution corresponds to a blood-pressure resolution of 4 mmHg and would probably be
regarded as too coarse to adequately zero the output.
Notch Filter
The notch filter is often used in biological recording. This is a special filter that is designed to
eliminate a single fundamental frequency and very little else. Notch filters at 50 or 60 Hz are
commonly used to eliminate line-frequency pickup.
Notch filters remove a narrow section of the power spectrum and are often used in conjunction with
frequency-domain analysis. Nevertheless, many experimenters prefer to record the data "as is" and
then to digitally remove the offending frequency component from the power spectrum.
In the time domain, a notch filter must be used with caution. If the recording bandwidth encompasses
the notch filter frequency, signals that include a harmonic at the notch frequency will be grossly
distorted, as shown in Figure . On the other hand, if the notch filter is in series with a low-pass or
high-pass filter that excludes the notch frequency, distortions will be prevented. For example, notch
filters are often used during electromyogram (EMG) recording in which the line-frequency pickup is
sometimes much larger than the signal. The 50 or 60 Hz notch filter is typically followed by a
300 Hz high-pass filter. The notch filter is required because the high-pass filter does not adequately
reject the 50 or 60 Hz hum.
Vue de la page 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 156 157

Commentaires sur ces manuels

Pas de commentaire