CMD300 CONTROL SOFTWARE Page 71
C
YBER
A
MP
380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
4)
CO
UPLING + <coupling>
or
CO
UPLING - <coupling>
or
CO
UPLING + <coupling> - <coupling>
This command sets the positive and/or negative input coupling for the selected channels. The
allowed values for <coupling> are:
G
ND,
D
C, 0.1, 1, 10, 30, 100 and 300.
5)
DEVICE <device number>
or
DEVICE OFF
This command specifies Axon expandable RS232 bus device numbers. By default, the
program does not use device numbers and hence assumes that the CyberAmp is the only
device on the RS232 bus on that COM port. If there are multiple units on the RS232 bus,
then each unit will need to be explicitly addressed by its unique device number. Once a
device number is used, it is remembered by the CMD300 program until device numbers are
disabled (DEVICE OFF) or until another device number is specified.
The <device number> may be in the range 0-9. The current device number may be reviewed
with the STATUS DEFAULTS command.
6)
FILTER <filter value>
or
F
ILTER
OF
F
This command sets the low-pass filter for the selected channels. To bypass the low-pass filter
use the command FILTER OFF. The permissible values for <filter value> are as follows:
2
4 40 400 4000
6 60 600 6000
8 80 800 8000
10 100 1000 10000
12 120 1200 12000
14 140 1400 14000
16 160 1600 16000
18 180 1800 18000
20 200 2000 20000
22 220 2200 22000
24 240 2400 24000
26 260 2600 26000
28 280 2800 28000
30 300 3000 30000
The letter "k" is a permissible shorthand for the whole multiples of 1000. (e.g., 10k is
equivalent to 10000).