Microsoft LPF-00004 Manuel de l'opérateur Page 84

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 157
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 83
Page 66 CMD300 CONTROL SOFTWARE
C
YBER
A
MP
380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
respecified unless the group is changed. The syntax of channel selection is flexible and allows lists
and ranges. For example:
CMD300 CHANNEL 1 'Selects only channel 1
CMD300 CHANNEL 1 2 5 'Selects channels 1,2 and 5
CMD300 CHANNEL 1-8 'Selects channels 1 thru 8
CMD300 CHANNEL 1-4 8 'Selects channels 1 thru 4 and 8
CMD300 CHANNEL ALL 'Selects all channels (1-8)
CMD300 CHANNEL ALL 2 'Selects all channels except 2
If more than one CyberAmp is being controlled, each will have a different Axon RS232 bus device
number (see the DEVICE command). CMD300 remembers the last device number that was used.
The parsing of the command line recognizes alpha character sequences as being distinct from numbers
and special characters, and abbreviations are recognized. Multiple commands may be entered on one
command line, and can be separated from each other by one or more spaces.
Note that most CMD300 commands have minimum abbreviations that can be used interchangeably
with the full commands. For example, the minimum abbreviation for the "STATUS" command is "S".
Likewise, most command modifiers also have minimum abbreviations. For example, the command
"STATUS DEFAULTS" could be entered as "S DEF". The minimum abbreviation for all commands
and modifiers is documented in the command reference section of this chapter.
Hence the command:
CMD300 C1G10C3-6G2
is the same as
CMD300 CHANNEL 1 GAIN 10 CHANNEL 3 - 6 GAIN 2
If CMD300 is run without any arguments then the program will display the status of the CyberAmp
and all channels. There is also an equivalent explicit "STATUS" command.
CMD300 STATUS
will produce the following output.
Program CMD300 Version 1.0
Copyright (c) Axon Instruments Inc. 1990. All rights reserved.
CYBERAMP 380 (Serial number 0001) (Microcode version 1.0)
#1 Pr=AI410 +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0mV
#2 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0mV
#3 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0mV
#4 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0mV
#5 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0mV
#6 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0mV
#7 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0mV
#8 Pr=None +=DC -=Gnd Gain=1(1,1) Filt=10k Notch=N Offset=+0mV
If the word "DEFAULTS" follows the explicit STATUS command, then the status of the program's
defaults is displayed.
Vue de la page 83
1 2 ... 79 80 81 82 83 84 85 86 87 88 89 ... 156 157

Commentaires sur ces manuels

Pas de commentaire