Microsoft LPF-00004 Manuel de l'opérateur Page 52

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 157
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 51
Page 34 PRINCIPLES OF OPERATION
C
YBER
A
MP
380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
Electrode Test
The electrode resistance is measured for several reasons. First, it establishes the basic continuity of
the electrode circuit. Sometimes electrodes or electrode leads break, causing the measured resistance
of previously low resistance electrodes to go very high and consequently resulting in no incoming
data. Second, it verifies that the electrode is making acceptable contact, as in surface electrodes. For
example, to achieve low-noise recording levels the electrode resistance may have to be less than
5 k. Third, the recording surface of the electrode can change. Insulation on the electrode can
develop cracks or other defects, leading to an increase in surface area and, in previously high
resistance electrodes, a decrease in resistance. Many researchers therefore keep a daily record of the
resistance of each electrode with respect to the reference electrode, or between electrodes in a
differential configuration.
When the Electrode Test is activated, a 1 V
p-p
(±0.5 V), 10 Hz square wave is connected to every
input via individual 1 M resistors. (Note that these are the same 1 M resistors that establish the
input resistance of the CyberAmp inputs.) The connections are shown in Figure 14.
The square wave current may not always produce a square wave response. Some electrodes may
polarize during the test, showing an apparent charging curve. This will not detract from a measure of
the ohmic resistance of the electrode if the response reaches an asymptotic peak value. Users may
prefer to inspect a simpler sinusoidal response, obtained by setting the lowpass filter to 12 Hz.
Figure 15 shows the simplified form of a highly filtered Electrode Test response.
R
e
e
R
INPUT
INPUT
DC
AC
V
IN
1 M
1 M
GND
GND
TEST
S1
NORMAL
10 Hz
1 V
p - p
DC
AC
GAIN
o
V
Figure 14. The connection of the Electrode Test circuit to the inputs of a CyberAmp channel. When switch S1 is
in the NORMAL position the 1 M resistors at the input provide the standard 1 M input resistance of the
CyberAmp. When S1 is in the TEST position, a 1 Vp-p square wave is applied to the 1 M resistors. The negative
input is grounded in this example, so that the only voltage that is recorded is Vin on the positive input. The
magnitude of Vin is described by Equations 1 and 2 in the text.
Vue de la page 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 156 157

Commentaires sur ces manuels

Pas de commentaire